• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRISMATIK DENTALCRAFT, INC. 3D SOFT SPLINT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRISMATIK DENTALCRAFT, INC. 3D SOFT SPLINT Back to Search Results
Device Problem Insufficient Information (3190)
Patient Problem Hypersensitivity/Allergic reaction (1907)
Event Type  Injury  
Manufacturer Narrative
The device has not been returned.If/when there is more information provided, a supplemental report will be submitted.
 
Event Description
It was reported that the it is unclear when the patient received the device, when the patient first used the device, or when the reaction occurred.Also noted was a "darkened lip line.".
 
Manufacturer Narrative
The device was returned, the investigation has been completed and the results are as follows: dhr results the dhr was reviewed for case #(b)(6) and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.The receiving router for lot# lg2743 indicated printer resin from keystone industries (supplier) passed inspection and met the specification.C of a (certificate of analysis) was also provided and confirmed the material was qualified for release stock product reviewed results no stock product was available for review since the device was fabricated per physician's prescription only.Investigation methods/results complaint investigator reviewed the returned device.Comfort3d bite splint was returned in the original case.The results were summarized: roughness - the flange was smooth; internal/external surfaces were smooth.Crack - no major crack was found.Delamination - layers were intact and did not separate.Discoloration - the device did not show any signs of discoloration.General cleanliness - the returned device appeared to be clean, free of debris or foreign particles.The returned device was visually inspected, and no defect or abnormality was observed.There was no evidence found to indicate that the reported issue was caused by the device itself.Root cause a root cause for this complaint cannot be explicitly determined.Ifu 12383 rev 3.0 (comfort3d bite splint instruction for use) states "brush and floss your teeth before use.Rinse mouth well with clean water before inserting the device.If a patient uses mouthwash, all traces of mouthwash should be removed by thoroughly rinsing out mouth with water.Rinse bite splint well with clean, cool water before and after use.Clean bite splint with clean, cool water only and let air dry." ifu provides warning "do not clean or soak in mouthwash; do not use denture cleanser, hot water, alcohol, hydrogen peroxide; do not place in direct sunlight".It is possible that reactions could be caused by mouthwash, toothpaste, or soaking material.However, the customer did not provide the information regarding how the patient handled and maintained the device.Rpt 012620 rev.1.0 (comfort3d bite splint verification), confirms that the resin material meets the acceptance criteria for biocompatibility (cytotoxicity, irritation, and sensitization) and mechanical properties (flexural strength, flexural modulus, sorption, solubility, and free monomer extraction).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3D SOFT SPLINT
Type of Device
3D SOFT SPLINT
Manufacturer (Section D)
PRISMATIK DENTALCRAFT, INC.
2144 michelson drive
irvine CA 92612
Manufacturer (Section G)
PRISMATIK DENTALCRAFT, INC.
2144 michelson drive
irvine CA 92612
Manufacturer Contact
herbert crane
2144 michelson drive
irvine, CA 92612
9495021907
MDR Report Key14454136
MDR Text Key293355462
Report Number3011649314-2022-00283
Device Sequence Number1
Product Code MQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/20/2022
Initial Date FDA Received05/20/2022
Supplement Dates Manufacturer Received09/27/2022
Supplement Dates FDA Received12/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexFemale
-
-