Catalog Number PEG-24-PULL-S
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abdominal Pain (1685); Laceration(s) (1946); Peritonitis (2252); Laceration(s) of Esophagus (2398); Post Operative Wound Infection (2446); Respiratory Failure (2484); Fluid Discharge (2686); Gastrointestinal Hemorrhage (4476); Skin Inflammation/ Irritation (4545)
Event Date 12/01/2008
Event Type  Injury  
Event Description
Cook endoscopy was notified of this event involving cook peg 24 percutaneous endoscopic gastrostomy set - pull via a clinical literature article.This article was published in 2017.Please see below for relevant excerpts of this article.This study was based on a retrospective analysis of a prospectively collected database of all patients with hnc (head and neck cancer) who underwent peg at the cancer institute of sao paulo from december 2008 to october 2013.The indication for peg placement was made by the treating oncologist or head and neck surgeon.The following complications occurred: respiratory failure (2), buried bumper syndrome (6), peritonitis (1), upper gi bleeding (1), minor bleeding (1), esophageal laceration (2), mallory - weiss syndrome (1), wound infection (15), peristomal leakage (5), abdominal pain (7), granulation tissue (7) and dermatitis (3).Major complications were resolved with endoscopy, surgical intervention, or prolonged intervention.Minor complications required only local and conservative treatment.Note: these events are still under review and investigation.Separate reports to further capture the different severities of complications may be submitted based on the outcome of the investigation.
 
Manufacturer Narrative
The investigation is on-going.A follow-up emdr will be provided within 30 days of submission of this report.Note: these events are still under review and investigation.Separate reports to further capture the different severities of complications may be submitted based on the outcome of the investigation.Literature review: comparison of the pull and introducer percutaneous endoscopic gastrostomy techniques in patients with head and neck cancer.United european gastroenterology journal 2017, vol.5(3) 365¿373 author(s) 2016 reprints and permissions: sagepub.Co.Uk/journalspermissions.Nav doi: 10.1177/2050640616662160 journals.Sagepub.Com/home/ueg.
 
Manufacturer Narrative
The investigation is on-going.A follow-up emdr will be provided within 30 days of submission of this report.
 
Event Description
The following was originally reported: "cook endoscopy was notified of this event involving cook peg 24 percutaneous endoscopic gastrostomy set - pull via a clinical literature article.This article was published in 2017.Please see below for relevant excerpts of this article.This study was based on a retrospective analysis of a prospectively collected database of all patients with hnc (head and neck cancer) who underwent peg at the cancer institute of sao paulo from december 2008 to october 2013.The indication for peg placement was made by the treating oncologist or head and neck surgeon.The following complications occurred.Respiratory failure (2) buried bumper syndrome (6) peritonitis (1) upper gi bleeding (1), minor bleeding (1) esophageal laceration (2), mallory - weiss syndrome (1), wound infection (15), peristomal leakage (5), abdominal pain (7), granulation tissue (7) and dermatitis (3) major complications were resolved with endoscopy, surgical intervention, or prolonged intervention.Minor complications required only local and conservative treatment.Note: these events are still under review and investigation.Separate reports to further capture the different severities of complications may be submitted based on the outcome of the investigation." the subject of this report is now "respiratory failure (2 patients).Surgical intervention was required." the other published aes will be captured in separate emdrs.
 
Manufacturer Narrative
Literature review: comparison of the pull and introducer percutaneous endoscopic gastrostomy techniques in patients with head and neck cancer.United european gastroenterology journal 2017, vol.5(3) 365¿373 author(s) 2016 reprints and permissions: sagepub.Co.Uk/journalspermissions.Nav doi: 10.1177/2050640616662160 journals.Sagepub.Com/home/ueg.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A review of the device history record could not be conducted because the lot number was not provided.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Respiratory failure is a known clinical complication.There is no evidence to suggest the report is due to a device failure.The ifu states the following potential complications: ¿potential complications associated with placement and use of a peg tube include, but are not limited to: bronchopulmonary aspiration and pneumonia, respiratory distress or airway obstruction, peritonitis or septic shock, colocutaneous, gastrocolocutaneous or small bowl fistula, gastric dilation, sigmoid intra-abdominal herniation and volvulus, persistent fistula following peg removal, esophageal injury, necrotizing fasciitis, candida cellulitis, improper placement of inability to place peg tube, tube dislodgement or migration, hemorrhage, and tumor metastasis." prior to distribution, all peg 24 percutaneous endoscopic gastrostomy set - pull are subjected to a visual and functional inspection to ensure device integrity.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEG 24 PERCUTANEOUS ENDOSCOPIC GASTROSTOMY SET - PULL
Type of Device
KNT, TUBE, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key14458497
MDR Text Key292199402
Report Number1037905-2022-00231
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberPEG-24-PULL-S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/22/2022
Initial Date FDA Received05/20/2022
Supplement Dates Manufacturer Received05/19/2022
05/19/2022
Supplement Dates FDA Received06/15/2022
07/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-