Model Number FOL0102
Device Problem Sharp Edges (4013)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/29/2022
Event Type  malfunction  
Event Description
It was reported that there was a design flaw in the foley statlock and this issue has been going on for the patient for a long time.Stated that there was a little post and it snapped on the top part of the product when it pulled from the leg bag it tore a hole in the tube and the content would leak everywhere.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that there was a design flaw in the foley statlock and this issue has been going on for the patient for a long time.Stated that there was a little post and it snapped on the top part of the product when it pulled from the leg bag it tore a hole in the tube and the content would leak everywhere.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿clamp spikes too long¿.It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labelling could not have prevented the reported failure.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL FOLEY SWIVEL SILICONE TRICOT 25BX
Type of Device
STATLOCK DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14462200
MDR Text Key292494994
Report Number1018233-2022-03978
Device Sequence Number1
Product Code EYJ
UDI-Device Identifier00801741076114
UDI-Public(01)00801741076114
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberFOL0102
Device Catalogue NumberFOL0102
Device Lot NumberJUFT2524
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/29/2022
Initial Date FDA Received05/20/2022
Supplement Dates Manufacturer Received07/12/2022
Supplement Dates FDA Received07/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-