Model Number N/A |
Device Problems
Break (1069); Positioning Failure (1158); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/27/2022 |
Event Type
malfunction
|
Event Description
|
This is filed to report the cable break and the steerable guide catheter (sgc) unable to move in the minus direction.It was reported during device preparation, that when applying the minus knob on the steerable guide catheter (sgc), a pop was heard, and the device would not respond.The physician suspected a cable break, and the device was exchanged for a new sgc.There was no clinically significant delay in the procedure and no patient involvement.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information and without the device to analyze, a cause of the suspected cable break could not be determined.The reported popping noise and device unable to move in the minus direction appears to be cascading effects of the cable break.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|