Model Number 72290038
Device Problem Failure to Shut Off (2939)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/30/2022
Event Type  malfunction  
Event Description
It was reported that during an arthroscopy, two werewolf coblation wands would not turn off after use and after removing from the joint.The only way to stop it was to press the coagulation button and cause a system error to show up on the werewolf console.The procedure was completed with a non-significant delay using the same device.No patient complications were reported.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
H10: h2: additional information on g2.D4, lot no.: the following are the lot numbers reported: 2073823 and 2081962.However, it is unknown which of the two failed.D4, exp.Date: expiration dates for the two lots reported: 25-may-2024 (lot 2073823) and 02-nov-2024 (lot 2081962).H4, mgf.Date: manufacturing dates for two lots reported: 25-may-2021 (lot 2073823) and 02-nov-2021 (lot 2081962).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEREWOLF FLOW 90 COBLATION WAND
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14470350
MDR Text Key292475452
Report Number3006524618-2022-00247
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00885556647462
UDI-Public00885556647462
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K183346
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72290038
Device Catalogue Number72290038
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/04/2022
Initial Date FDA Received05/23/2022
Supplement Dates Manufacturer Received11/28/2022
Supplement Dates FDA Received11/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-