• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US MODULAR CATHCART BALL 54MM OD; HEMI HIP IMPLANT : HIP ACETABULAR METAL HEMI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US MODULAR CATHCART BALL 54MM OD; HEMI HIP IMPLANT : HIP ACETABULAR METAL HEMI Back to Search Results
Model Number 1363-54-000
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Date 05/09/2022
Event Type  Injury  
Event Description
Prosthesis was removed due to infection.No surgical delay.Doi: (b)(6) 2022.Dor: (b)(6) 2022.Affected side: left hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR CATHCART BALL 54MM OD
Type of Device
HEMI HIP IMPLANT : HIP ACETABULAR METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14471338
MDR Text Key292388290
Report Number1818910-2022-09481
Device Sequence Number1
Product Code LZY
UDI-Device Identifier10603295032861
UDI-Public10603295032861
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K903084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number1363-54-000
Device Catalogue Number136354000
Device Lot NumberJ87A87
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CORAIL REVISION STEM HO 14; TAPERED SPACER ARTICUL/EZE +5
Patient Outcome(s) Required Intervention;
Patient Age73 YR
Patient SexMale
-
-