Reportable based on device analysis completed on 03may2022.It was reported that the device could not track on diagnostic catheter.Vascular access was obtained via the femoral artery.The target lesion was located in the gastrointestinal area.A 130cm direxion transend-14 system was selected for use.The device was hydrated 15 seconds prior to removal of the hoop.Upon introduction, the device could not track on diagnostic catheter.The procedure was completed with another of same device.No complications were reported and patient was stable post procedure.However, device analysis revealed that the device was fractured 66.7cm from the hub.
|
Initial reporter address 1: (b)(6).Device evaluated by manufacturer: the device was returned for analysis.The device was inspected for any damage or irregularities.The direxion was received with the guidewire still fully inserted.The direxion was hydrated, and the guidewire was removed with extreme care due to the kinks and fracture.The device was inspected for kinks or damage, kinks were noted 4.5 cm, 4.7 cm, and 6.4 cm from the tip.A fracture was confirmed 66.7 cm from the hub.The shaft was not completely separated, the inner liner was still intact.The fracture looked consistent with a bending and tensile force breaking.The catheter had signs of stretching along the entire length and measured 61.5 cm to the tip.The device was not introduced into a guide catheter due to the kinks and fracture.Device analysis determined the condition of the returned device was consistent with the reported complaint of difficult to insert.
|