Model Number 505DM24
Device Problem Incomplete Coaptation (2507)
Patient Problem Mitral Valve Insufficiency/ Regurgitation (4451)
Event Date 04/28/2022
Event Type  Injury  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during the implant of this 24mm mitral mechanical valve, it was explanted and replaced with a 27mm non-medtronic bioprosthetic valve.The reason for the replacement was reported to be due to the two leaflets being unable to fully close resulting in severe regurgitation.It was also stated that upon opening, one of the valve leaflets was observed to be "a little sunken".The valve motion was not tested with the blue actuator.Tt was reported leaflet movement was normal and no impingement was noted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN PIVOT AP360 HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
Manufacturer (Section G)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14479047
MDR Text Key292489183
Report Number3008592544-2022-00024
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P990046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number505DM24
Device Catalogue Number505DM24
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2022
Date Device Manufactured08/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age79 YR
Patient SexFemale
-
-