Catalog Number V041441000000 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
As a result of this malfunction, the potential for surgical intervention exists to preclude injury or illness that would necessitate medical or surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event it is reported that eddy tip broke during use.Fragment removed, no injury occurred.
|
|
Manufacturer Narrative
|
Returned product was investigated, no smooth breakage, melted.Breakage due to thermal influence.A dhr review was conducted with no discrepancies noted.
|
|