MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number AFAPRO28 |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
|
Patient Problems
Hemorrhage/Bleeding (1888); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 04/25/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryoablation procedure, the left atrial pressure was elevated.Gross blood was noted in endotracheal tube (ett) and oxygen circuit.The procedure was immediately stopped and protamine was given to reverse the effects of heparin.The case was aborted under general anesthesia.Bronchoscopy was performed after the procedure.Right pulmonary artery bleed was noted, the patient was transferred to the intensive care unit (icu) with iced saline lavage to halt the bleeding. no further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the data files were returned and analyzed.The data files showed that at least nine applications were performed with a non-returned afapro28 balloon catheter with lot 15619 without any issue on the date of the event.In conclusion, clinical issues (left atrial pressure was elevated, hemorrhage) occurred during the procedure.There is no indication of relation of adverse events to the performance and malfunction of the product.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|