The customer, a (b)(6) authorized distributor, reported that the freedom driver exhibited a fault alarm and the fill volume and cardiac output values were very low while supporting a patient.There was no reported adverse patient impact.The customer also reported that the patient was subsequently switched to a backup driver.
|
Device history record (dhr) review confirmed that freedom driver s/n 5176 was serviced and passed all functional testing prior to being released to finished goods.Alarm history and patient data file review found one new alarm indicating secondary motor voltage too high.There was no evidence of secondary motor engagement, therefore, fault code was likely produced in house during data extraction.Visual inspection of external and internal components found no abnormalities.Freedom driver passed all sections of functional testing at incoming inspection.An additional 24-hour observation run was performed, resulting in the driver's cardiac output occasionally dropping out of specification.No alarms were produced.A known functional piston cylinder assembly was installed as a replacement part for testing purposes.Driver passed all sections of functional testing after a second 24-hour observation run.No alarms were produced, no abnormalities with cardiac output were observed, and driver functioned as intended.Complaint could not be replicated as reported, however, pca was determined to be faulty contributing to low output values.Failure investigation for this complaint confirmed the reported issue.Although the alarm could not be replicated during functional and observational testing, it was determined that the low output values were due to a faulty pca.Failure investigation identified no other test failures or damage that could have contributed to the complaint.Patient was switched to a backup driver with no reported adverse impact.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
|
The customer, a syncardia authorized distributor, reported that the freedom driver exhibited a fault alarm and the fill volume and cardiac output values were very low while supporting a patient.There was no reported adverse patient impact.The customer also reported that the patient was subsequently switched to a backup driver.
|