MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE HIGH FLOW INSUFFLATOR; INSUFFLATOR, LAPAROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE HIGH FLOW INSUFFLATOR; INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Model Number 0620040600
Device Problems Pressure Problem (3012); Therapeutic or Diagnostic Output Failure (3023)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2022
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was loss of insufflation during procedure.
 
Manufacturer Narrative
The device manufacturer date is not known.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: loss of insufflation.Probable root cause: pressure sensor malfunction / out of calibration; software malfunction; use error; system design; unwanted movement of internal components / wiring; power button inadvertently turned off; tube set/gas supply inadvertently detached/loose; loss of power; pressure button does not disengage; electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge; hpu or lpu assembly malfunction; leaks from internal connections or seals; ppv failure; manufacturing/ service error.The reported failure mode will be monitored for future reoccurrence.H3 other text: 81.
 
Event Description
It was reported that there was loss of insufflation during procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG; PNEUMO SURE HIGH FLOW INSUFFLATOR
Type of Device
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key14516438
MDR Text Key296508393
Report Number0002936485-2022-00269
Device Sequence Number1
Product Code HIF
UDI-Device Identifier07613327063219
UDI-Public07613327063219
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0620040600
Device Catalogue Number0620040600
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-