BOSTON SCIENTIFIC CORPORATION CAROTID WALLSTENT; STENT, CAROTID Back to Search Results
Device Problem Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/07/2022
Event Type  Injury  
Event Description
It was reported that the overlapping stents migrated and a new stent was placed in between.The 90% stenosed target lesion was located in the severely tortuous and moderately calcified vein.A 10x31mm carotid wallstent was selected for use.During the procedure, dilatation was performed with a 3.0x20mm sterling balloon and the 10x31mm carotid wallstent was deployed.The distal vessel with the deployed stent was severely tortuous and an 8x21mm carotid wallstent was then placed to overlap the first carotid wallstent by 5mm.Contrast with 5x20mm sterling balloon was then used for post dilation.Afterwards, it was noted that the stents migrated and were no longer overlapped.A third carotid wallstent was used, and it was placed in between the two placed stents to complete the procedure.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14518286
MDR Text Key292795878
Report Number2134265-2022-05735
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/07/2022
Initial Date FDA Received05/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-