• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL SA ASEPTIC BATTERY HOUSING; SURGICAL POWER TOOL BATTERY-EXCHANGE STERILITY COVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL SA ASEPTIC BATTERY HOUSING; SURGICAL POWER TOOL BATTERY-EXCHANGE STERILITY COVER Back to Search Results
Model Number N/A
Device Problems Break (1069); Fracture (1260)
Patient Problem Bone Fracture(s) (1870)
Event Date 05/11/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the battery box opened during the procedure and the cover was torn off during a femoral cut.The handpiece stopped suddenly when the housing opened and it caused a calcaneus fracture.The surgery was prolonged by 15 minutes because they needed a new power tool and the use of cerclage for the fracture.Attempts have been made and all additional information received has been included in this report.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The product was return for investigation.Upon receipt, the lid's hinge was broken.Also, the box had a scratch.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASEPTIC BATTERY HOUSING
Type of Device
SURGICAL POWER TOOL BATTERY-EXCHANGE STERILITY COVER
Manufacturer (Section D)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ  CH-1228
Manufacturer (Section G)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ   CH-1228
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key14520481
MDR Text Key292788424
Report Number0008031000-2022-00013
Device Sequence Number1
Product Code MOQ
UDI-Device Identifier889024581746
UDI-Public(01)889024581746(11)210929(10)5015326
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number89-8521-470-40
Device Lot Number5015326
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-