Model Number N/A |
Device Problems
Break (1069); Fracture (1260)
|
Patient Problem
Bone Fracture(s) (1870)
|
Event Date 05/11/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the battery box opened during the procedure and the cover was torn off during a femoral cut.The handpiece stopped suddenly when the housing opened and it caused a calcaneus fracture.The surgery was prolonged by 15 minutes because they needed a new power tool and the use of cerclage for the fracture.Attempts have been made and all additional information received has been included in this report.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The product was return for investigation.Upon receipt, the lid's hinge was broken.Also, the box had a scratch.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|