Model Number 990063-020
Device Problems Entrapment of Device (1212); Device Handling Problem (3265)
Patient Problems Chest Pain (1776); Hemorrhage/Bleeding (1888)
Event Date 05/18/2022
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that at the end of a completed cryo ablation procedure, the mapping catheter could not be removed from the left superior pulmonary vein (lspv) which had multiple side branches.The mapping catheter seemed to be "hooked" on a structure within the patient's anatomy.After multiple different maneuvers, the mapping catheter was able to be removed.A computerized tomography (ct) was performed and a small hemorrhage was observed in the lower lobe of the lung.The physician suspected that the hemorrhage was due to the maneuvering of the mapping catheter.The patient was hospitalized with mild thoracic pressure.The following day, the patient was stable without any symptoms of a hemorrhage. no further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the 990063-020 mapping catheter with lot number 223937855 was returned and analyzed.The mapping catheter was received inserted inside the balloon catheter, unable to retract the mapping catheter.Visual inspection of the loop segment area showed the loop shape was deformed and out of plane.The loop was kinked and ribbed near the electrodes.The user may have experienced difficulties introducing the mapping catheter into the balloon catheter.In conclusion, the reported maneuvering issues were confirmed through testing.The mapping catheter failed the returned product inspection due to the kinked and ribbed loop.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER - 20 MM
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14523517
MDR Text Key292822433
Report Number9617601-2022-00074
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K102588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/14/2024
Device Model Number990063-020
Device Catalogue Number990063-020
Device Lot Number223937855
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/19/2022
Initial Date FDA Received05/27/2022
Supplement Dates Manufacturer Received06/07/2022
Supplement Dates FDA Received07/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient SexMale
-
-