• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 56MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 56MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74120156
Device Problems Degraded (1153); Biocompatibility (2886)
Patient Problems Cyst(s) (1800); Failure of Implant (1924); Arthralgia (2355); Metal Related Pathology (4530)
Event Date 08/02/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, after a bhr resurfacing construct had been implanted on the patient¿s left hip on (b)(6) 2010, the patient experienced extensive metallosis, joint pain, large anterior fluid cyst and damage to antero-superior region of the acetabular implant.A revision surgery was performed on (b)(6) 2021 to treat this adverse event.During this procedure, an extensive metallosis staining of the soft tissue around the hip was visualized and carefully debrided.The bhr resurfacing head was removed upon performing a femoral neck cut.The acetabular cup presented an anterior superior wear and a defect in the surface of the cup, making it necessary to revise this component as well.The hip was converted into a full thr system.The patient was transferred to the recovery room in stable condition.
 
Manufacturer Narrative
H3, h6: it was reported that a let hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the acetabular cup and femoral head was performed using batch numbers to evaluate patterns of repeated failures or defects over the lifetime of the product and using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware date no other similar complaints were identified to involve this batch.Other similar complaints were identified for the part number and the reported failure mode.This will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions required.The available medical documents were reviewed.Based on the reported symptoms it cannot be concluded that the events/clinical reactions (pain, extensive metallosis, fluid collection, and damage to the acetabular cup) were associated with a mal-performance of the implant.The patient impact cannot be determined at this time.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Event Description
It was reported that, after a bhr resurfacing construct had been implanted on the patient¿s left hip on (b)(6) 2010, the patient experienced extensive metallosis, joint pain, large anterior fluid cyst and damage to antero-superior region of the acetabular implant.A revision surgery was performed on (b)(6) 2021 to treat this adverse event.During this procedure, an extensive metallosis staining of the soft tissue around the hip was visualized and carefully debrided.The acetabular cup presented an anterior superior wear and a defect in the surface of the cup, making it necessary to revise this component.The hip was converted into a full thr system.The patient was transferred to the recovery room in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 56MM W/ IMPTR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14527516
MDR Text Key293301459
Report Number3005975929-2022-00354
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010502599
UDI-Public03596010502599
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/30/2015
Device Model Number74120156
Device Catalogue Number74120156
Device Lot Number10GW28980
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/05/2022
Initial Date FDA Received05/27/2022
Supplement Dates Manufacturer Received08/01/2022
Supplement Dates FDA Received08/07/2022
Date Device Manufactured07/31/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other; Hospitalization;
Patient Age64 YR
Patient SexFemale
-
-