• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
One unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.The device is out of calibration and the pcb is damaged which won't allow it to be adjusted.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.
 
Event Description
It was reported that the lcd display fluctuates on the device.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14547972
MDR Text Key293351922
Report Number3012307300-2022-10111
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/05/2022
Initial Date FDA Received05/31/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-