SMITH & NEPHEW, INC. LGN INTRA ALIGNMENT GUIDE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71434330 |
Device Problems
Detachment of Device or Device Component (2907); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/11/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that during a tka surgery the lgn intra alignment guide locking piece was missing.The guide could not be locked to the im rod.The procedure was completed, without delay, using a s+n back-up device.Patient was not harmed as consequence of this problem.
|
|
Manufacturer Narrative
|
H3, h6.The associated device was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.The device shows signs of extensive use.A functional evaluation of the returned device could not confirm the stated failure mode.Even though the device was worn, it was still found to be functional.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.The contribution of the device to the reported event could not be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|