Catalog Number 223.591
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in portugal as follows: it was reported on an unknown date that when the plate was delivered, it was found to be bent.No other information was provided.There was no patient consequence reported.This report is for one (1) lcp 3.5 9ho l124 sst.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6: part # 223.591.Lot # 367p302.Manufacturing site: jabil grenchen.Release to warehouse date: september 2, 2021.A manufacturing record evaluation was performed for the finished device lot and no non-conformance was identified.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photo.Visual analysis of the photo revealed that the lcp 3.5 9ho l124 sst was bent.While no definitive root cause can be established, it is most likely due that the device was damage during shipping/transportation process.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the lcp 3.5 9ho l124 sst.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: b5: subsequent follow-up with the customer, additional information was received regarding the event.It was reported that the event occurred when the plate was replaced not during a surgical procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCP 3.5 9HO L124 SST
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK GRENCHEN (CH)
solothurnstrasse 186
grenchen 2540
SZ   2540
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14564950
MDR Text Key300874284
Report Number8030965-2022-03593
Device Sequence Number1
Product Code HRS
UDI-Device Identifier07611819241091
UDI-Public(01)07611819241091
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K082807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number223.591
Device Lot Number367P302
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/19/2022
Initial Date FDA Received06/01/2022
Supplement Dates Manufacturer Received06/09/2022
07/07/2022
Supplement Dates FDA Received06/16/2022
07/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/02/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-