• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL CLARIA MRI CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL CLARIA MRI CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTMA2D1
Device Problems Unable to Obtain Readings (1516); Missing Test Results (3267)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/19/2022
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: 383074 lead implanted: (b)(6) 2020.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a follow up appointment the cardiac resynchronization therapy defibrillator (crt-d) showed a longevity gray bar with question marks for a longevity estimate.Upon review of the device file the correct battery longevity was available and normal. the device is still in use. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the battery measurement was not available.Analysis of the device memory indicated an issue with missing/invalid diagnostic data.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARIA MRI CRT-D SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14565505
MDR Text Key293086506
Report Number9614453-2022-01590
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/14/2022
Device Model NumberDTMA2D1
Device Catalogue NumberDTMA2D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/01/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4470-52 LEAD, 693065 LEAD
Patient Age87 YR
-
-