• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Failure to Auto Stop (2938)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2022
Event Type  malfunction  
Event Description
An olympus employee reported the insufflation unit did not stop insufflation automatically.This is a loaner device and was not in use at a facility.No patient or procedure was involved in the event.
 
Manufacturer Narrative
The suspect device was sent to an olympus service center for evaluation.During inspection and testing, the reported event was not confirmed.The inspection found nothing out of standard.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been nearly 8 years since the subject device was manufactured.The device met all specifications at the time of shipment.Based on the results of the investigation, and since no device malfunction was confirmed during evaluation, the definitive root cause of the reported issue could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14566102
MDR Text Key301012886
Report Number8010047-2022-09194
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/05/2022
Initial Date FDA Received06/01/2022
Supplement Dates Manufacturer Received06/13/2022
Supplement Dates FDA Received07/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-