Device Problem Insufficient Information (3190)
Patient Problem Hypersensitivity/Allergic reaction (1907)
Event Date 03/08/2022
Event Type  Injury  
Event Description
It was reported that the patient had a reaction to the comfort hard soft splint.It is unclear when the patient received the device.However, the device was first used (b)(6) 2022 with the reaction occurring that same day to the gingival tissue.The reaction resolved within 48 hours after the patient discontinued its use on (b)(6) 2022.
 
Manufacturer Narrative
The device has not been returned.If/when there is more information provided, a supplemental report will be submitted.
 
Manufacturer Narrative
The device has not been returned.However, the non-visual device evaluation has been completed and the results are as follows: dhr results the dhr was reviewed and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.Supplier (erkodent) reviewed the associated material lot and confirmed no material defects or changes could be detected.No manufacturing deviations or abnormalities are know.E-pro 4.0 - 11761112 (erkoloc-pro) was manufactured from (b)(6), 2021 and was assigned an expiration of (b)(6) 2024.Stock product reviewed results no stock product was available for review since the device was fabricated per physician's prescription only.Investigation methods/results no device has been returned from the customer and no picture was provided by the customer to review.However, the non-visual device investigation has been completed.A root cause for this complaint cannot be explicitly determined.No device has been returned for review.Ifu 9091 rev 5.0 (comfort h/s bite splint instruction for use) states "brush and floss your teeth before use.Rinse mouth well with clean water before inserting the device.If patient uses mouthwash, all traces of mouthwash should be removed by thoroughly rinsing out mouth with water.Rinse bite splint well with clean, cool water before and after use.Clean bite splint with clean, cool water only and let air dry." ifu provides warning "do not clean or soak in mouthwash; do not use denture cleanser, hot water, alcohol, hydrogen peroxide; do not place in direct sunlight".It is possible that reactions could be caused by mouthwash, toothpaste, or soaking material.This complaint will be kept on record for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMFORT HARD SOFT SPLINT
Type of Device
THERMOFORM MOUTHGAURD
Manufacturer (Section D)
PRISMATIK DENTALCRAFT, INC.
2144 michelson drive
irvine CA 92612
Manufacturer (Section G)
PRISMATIK DENTALCRAFT, INC.
2144 michelson drive
irvine CA 92612
Manufacturer Contact
herbert crane
2144 michelson drive
irvine, CA 92612
9495021907
MDR Report Key14566156
MDR Text Key293092005
Report Number3011649314-2022-00311
Device Sequence Number1
Product Code MQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121365
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 06/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/28/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
Patient RaceWhite
-
-