ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG Back to Search Results
Model Number 6662
Device Problem Wireless Communication Problem (3283)
Patient Problems Failure of Implant (1924); Inadequate Pain Relief (2388)
Event Date 05/17/2022
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
It was reported that, following radiation therapy, the patient's ipg was unable to communicate with external devices, deeming the ipg inoperable.In turn, surgical intervention was undertaken wherein the ipg was explanted and replaced, resolving the issue.
 
Manufacturer Narrative
The reported observation ¿ipg inoperable¿ was not confirmed or duplicated during analysis.The root cause of the observation was not ascertained.The ipg diagnostic logs showed the device appeared to be functioning normally on the observation date and was controlled/programmed after the event date.The device exhibited normal functionality during analysis which included communication testing with a clinician programmer.No programmer logs were returned from the field for review.The device was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR
Type of Device
DBS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key14569154
MDR Text Key293137017
Report Number1627487-2022-03065
Device Sequence Number1
Product Code MHY
UDI-Device Identifier05415067020260
UDI-Public05415067020260
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
P140009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/23/2021
Device Model Number6662
Device Catalogue Number6662
Device Lot Number7096396
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/24/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-