Model Number DSX500T11C
Device Problem Degraded (1153)
Patient Problems Angina (1710); Dyspnea (1816)
Event Date 01/11/2022
Event Type  malfunction  
Event Description
The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient alleged visualization of black particles.Patient alleged difficulty in breathing & chest pain.There is no allegation of serious or permanent harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer previously received information alleging an issue related to a cpap device's sound abatement foam.The patient alleged visualization of black particles.Patient alleged difficulty in breathing & chest pain.There is no allegation of serious or permanent harm or injury the device was returned to the manufacturer's product investigation laboratory for investigation.An internal visual inspection was completed by the manufacturer and found no evidence of contamination.The manufacturer found no evidence of sound abatement foam degradation.The device's downloaded event log was reviewed by the manufacturer and found 1 instance of e-4 on the error log.The manufacturer confirmed there was no evidence of sound abatement foam degradation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key14569548
MDR Text Key293122407
Report Number2518422-2022-27135
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 02/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500T11C
Device Catalogue NumberDSX500T11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/12/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES 88058
Patient Sequence Number1
-
-