As reported by the field, during a mechanical thrombectomy procedure, a 5x33 embotrap iii revascularization device (et007533, lot unknown) became ¿stuck¿ due to there being too much forward tension on the aspiration catheter pushing the device into the vessel wall.This was naturally resolved by removing that tension.No further info is available.
|
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Lot: the product lot number was not reported.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: as reported by the field, during a mechanical thrombectomy procedure, a 5x33 embotrap iii revascularization device (et007533, lot unknown) became ¿stuck¿ due to there being too much forward tension on the aspiration catheter pushing the device into the vessel wall.This was naturally resolved by removing that tension.No further info is available.The device was discarded; therefore, no further investigation can be performed.The lot number is not known; therefore, a device history record review cannot be completed.Withdrawal difficulty from vessel is a potential complication associated with the use of the embotrap iii revascularization device in endovascular mechanical thrombectomy.The embotrap instructions for use (ifu) warns the user to never withdraw the device against significant resistance.Assess the cause of resistance using fluoroscopy and if necessary, advance the microcatheter over the device to resheath or partially resheath to aid withdrawal.With the amount of information available and without the device for analysis, it is not possible to draw a conclusion between the device and the reported event.However, there are clinical, procedural factors, including vessel characteristics, severe tortuosity, clot burden/characteristics, device interaction, device selection, and operator technique that may have contributed to the event.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
|