MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Model Number ER320
Device Problems Failure to Form Staple (2579); Failure to Fire (2610); Firing Problem (4011)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #: unk.No lot or batch number was provided therefore a device history could not be done.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during a laparoscopic cholecystectomy they had to squeeze the handle several times to load the clip into the jaws.It then didn't fire and close properly resulting in a deformed clip.A new device was used to complete the case with no patient consequences.
 
Manufacturer Narrative
(b)(4).Date sent: 6/27/2022 investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the er320 device was returned with a clip in the jaws and no apparent damage.The clip was removed in order to inspect the jaws and they were found with no damage.The device was tested for functionality.During the analysis, the device was noted with fluids.The device was functional tested, and it fed, retained and formed the remaining 19 clip as intended.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
MDR Report Key14577546
MDR Text Key293250217
Report Number3005075853-2022-03563
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public10705036012580
Combination Product (y/n)N
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberER320
Device Catalogue NumberER320
Device Lot NumberV7016D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received06/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-