According to the reporter, during shavering a polyp in a hysteroscopy procedure, there was no pathology sucked in the device.They stopped the procedure and tested the vacuum (different settings were tested (window lock, full vacuum)).It looked like the shaver was clogged as it would not suck up any liquid.It was also reported that the procedure was not completed as there was no other blades available but there was no patient injury.
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit showed signs of interior/exterior contamination.Functionally, inspection of the laser markings on the inner and outer blade assemblies indicated that they are in the correct location.The inner blade rotated freely, and the blade could be placed into window locked manually.The fluid flow during window lock was measured under suction by connecting the device to a handpiece and controller.The device was unable to remove fluid during testing.Further investigation found that the innertube was blocked by foreign material.Functionality was restored by removing the blockage.It was reported that there was an adverse event without an identified device or use problem, the device was clogged, and had no suction.A potentially related device issue were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|