Catalog Number 136561100 |
Device Problems
Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Cyst(s) (1800); Fall (1848); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Inflammation (1932); Pain (1994); Discomfort (2330); Metal Related Pathology (4530); Swelling/ Edema (4577)
|
Event Date 04/13/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Initial reporter occupation: consumer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Revision completed due to patient developing a large pseudotumor.Patient started to complain of pain and swell.Mrs showed the pseudotumor.All components were removed.Doi: 2010, dor: (b)(6) 2022, affected side: right hip.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary = > update 17 jun 2022 no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot = > a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: update 15-july-2022.No device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: a2 (age), a4, b5, b6, b7, h6 (clinical codes).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: e3.E3 initial reporter occupation: lawyer.
|
|
Event Description
|
In addition to what previously alleged, patient alleges discomfort and fall.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H6: appropriate term / code not available (e2402) is being utilized to capture hypersensitivity/allergic reaction.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Litigation and medical records received.Litigation records has no new allegation reported.After review of medical records no revision notes were reported.In addition to what was previously alleged, the patient alleged metallosis.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot = > a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
|
|
Event Description
|
Litigation records received alleged heavy metal poisoning.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 (clinical codes: metal related pathology (e1618) used to capture blood heavy metal increased and metal poisoning) e3 initial reporter occupation: lawyer if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot : a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
|
|
Manufacturer Narrative
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Medical records were received and stated that on (b)(6) 2022, patient had a revision for right total hip replacement due to metallosis and localized fluid collection.Operative reported a darkish staining of metallic wear debris noted in anterior portion of acetabular device.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5, h6 health effect - clinical code if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Medical records received.Patient was revised due to the following pathologic findings: cyst, corrosion; metallosis.There was metallic staining at the base of the trunnion.
|
|
Search Alerts/Recalls
|