MEDTRONIC HEART VALVES DIVISION EVOLUT PRO PLUS DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number D-EVPROP2329US |
Device Problems
Failure to Advance (2524); Difficult to Advance (2920)
|
Patient Problem
Vascular Dissection (3160)
|
Event Date 05/25/2022 |
Event Type
Injury
|
Event Description
|
Medtronic received information that during implant of this transcatheter bioprosthetic valve, the delivery catheter system (dcs) was unable to advance.Several attempts were made.Re-dilation of the artery had to be performed in order to advance the dcs.After the procedure was completed, the bleeding continued.Surgery was called in to repair the right femoral artery.Per the physician, the transition on the nose cone of the dcs was poor, which resulted in the issues.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information was received that the minimum access vessel diameter was 5.9 mm and tortuous iliac arteries were noted prior to implant.It was reported that the bleeding was caused by a dissection.Per the physician, the dissection was caused by the delivery catheter system (dcs).A surgical cutdown was performed and the vessel was sutured successfully.A1 and h6 fdd code was corrected.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Updated data: conclusion: no device was returned for analysis and no procedural images were submitted for review.A device history record review was performed on the device and there were no correlations or issues identified regarding manufacturing.The device was manufactured per approved and released manufacturing processes and the device met all applicable manufacturing specifications prior to release for distribution.As no procedural images were provided, and no images of the reported transition on the nose cone were provided, an assignable root cause of the vascular complication cannot be determined and the relationship to the device could not be established.There was no information to suggest a device malfunction or a failure to meet manufacturing specifications was related to these events.Vascular access related complications, such as bleeding and dissection, are a known potential adverse patient effect per the device system instructions for use, and are typically related to patient factors (anatomy, comorbidities, etc.), and/or procedural effects (sheath used, user technique, puncture cut location, etc.).Per the physician, the transition on the nose cone of the delivery catheter system (dcs) was poor, which resulted in the issues.Per the physician, the dissection was caused by the dcs.Difficulties advancing the device through the access vessel is known to be related to factors such as patient anatomy and physician technique, including guidewire and introducer sheath selection.In this case, tortuous iliac arteries were noted prior to implant.This indicates that the probable cause of the advancement difficulties was patient anatomy, but this cannot be confirmed with the limited information available.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|