• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KINETIC CONCEPTS, INC. ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM; OMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KINETIC CONCEPTS, INC. ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM; OMP Back to Search Results
Model Number WNDARM
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Bacterial Infection (1735)
Event Date 05/05/2022
Event Type  Injury  
Event Description
(b)(6) 2022, the following information was reported to kci by the patient: on (b)(6) 2022, the activ.A.C.¿ ion progress¿ remote therapy monitoring system was placed on hold.On (b)(6) 2022, she was prescribed antibiotics allegedly for two types of infection.No additional information was provided.A device evaluation of the activ.A.C.¿ ion progress¿ remote therapy monitoring system is pending completion.On (b)(6) 2022, a device evaluation was completed by kci quality engineering.On (b)(6) 2022, the device was tested per quality control procedure by kci service center, and the unit passed the quality control checks and met specifications.On (b)(6) 2022, the device was placed with the patient.On (b)(6) 2022, the device was tested per quality control procedure by kci service center, and the unit passed the quality control checks and met specifications.Inspection and testing of the device did not reveal any evidence of an operational malfunction with the unit.
 
Manufacturer Narrative
Based on the information provided, it cannot be determined that the alleged bacterial infection requiring antibiotics is related to the activ.A.C.¿ ion progress¿ remote therapy monitoring system.The patient has a history of complications.Multiple unsuccessful attempts have been made to obtain additional clinical information.The device met specifications before and after patient placement.Device labeling, available in print and online, states: wound infection: call your doctor or nurse right away if you think your wound is infected or if the following symptoms develop or worsen: you have a fever.Your wound is sore, red or swollen.Your skin itches or you have a rash or redness around the wound.The area around the wound feels very warm.You have pus or a bad smell coming from the wound.Infected wounds: infected wounds should be monitored closely and may require more frequent dressing changes than noninfected wounds, dependent upon factors such as wound conditions, treatment goals.Refer to dressing application instructions (found in v.A.C.® dressing cartons) for details regarding dressing change frequency.As with any wound treatment, clinicians and patients / caregivers should frequently monitor the patient's wound, periwound tissue and exudate for signs of infection, worsening infection, or other complications.Some signs of infection are fever, tenderness, redness, swelling, itching, rash, increased warmth in the wound or periwound area, purulent discharge or strong odor.Infection can be serious, and can lead to complications such as pain, discomfort, fever, gangrene, toxic shock, septic shock and/or fatal injury.Some signs or complications of systemic infection are nausea, vomiting, diarrhea, headache, dizziness, fainting, sore throat with swelling of the mucus membranes, disorientation, high fever, refractory and/or orthostatic hypotension or erythroderma (a sunburn-like rash).If there are any signs of the onset of systemic infection or advancing infection at the wound site, contact the treating physician immediately to determine if v.A.C.® therapy should be discontinued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM
Type of Device
OMP
Manufacturer (Section D)
KINETIC CONCEPTS, INC.
san antonio TX 78249
Manufacturer Contact
steven jackson
6203 farinon drive
san antonio, TX 78249
2102556438
MDR Report Key14594920
MDR Text Key293289318
Report Number3009897021-2022-00105
Device Sequence Number1
Product Code OMP
UDI-Device Identifier00849554005600
UDI-Public0100849554005600
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201571
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberWNDARM
Device Catalogue Number420095
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2022
Date Manufacturer Received05/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age52 YR
Patient SexFemale
-
-