Model Number 1MTEC30
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2022
Event Type  malfunction  
Manufacturer Narrative
Weight and ethnicity: unknown/ asked but not available.If implanted, give date: not applicable, as this is not an implantable device.If explanted, give date: not applicable, as this is not an implantable device.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Device manufacture date: unknown as lot number was not provided.Attempts have been made to obtain missing information; however, the account did not provide the information.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that the intraocular lens (iol) failed to align with the plunger.Additional information received clarified the event and reported that the lens cartridge tip broke upon insertion into patient's right eye.There was no patient injury and no other intervention was required.Another johnson & johnson lens of same model and diopter was used as the replacement.No other information was available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNFOLDER PLATINUM 1 SERIES
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14598768
MDR Text Key301030648
Report Number3012236936-2022-01461
Device Sequence Number1
Product Code KYB
UDI-Device Identifier05050474540323
UDI-Public(01)05050474540323(10)UNKNOWN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081545
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1MTEC30
Device Catalogue Number1MTEC30
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ZCB00 LENS SN:(B)(4)
Patient SexMale
-
-