H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, a uni-knee insert revision/poly swap was performed due to insert dislocation ¿a few months¿ post uni-knee implantation.The requested medical documentation has not been received as of the date of this medical investigation and the current patient status is unknown.Without the requested clinical documentation, the root cause of the reported events could not be further assessed.The patient impact beyond the reported revision/poly swap could not be determined.No further medical assessment could be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the instructions for use for this device revealed that dislocation has been identified as as result of trauma, improper implant selection, position or fixation and/or migration of the components.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|