Catalog Number L1-CW-100V |
Device Problems
Device Emits Odor (1425); Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that during the use of the product, a burnt smell from detected.There was no patient injury, and additional information is not available.
|
|
Manufacturer Narrative
|
Manufacturing device history record review was not performed because manufacturing records are retained by the supplier at their manufacturing site and are not readily available for review.The device was returned with no physical damaged was found on the device.The reported problem was duplicated, as the product operated it smelled burnt.The root cause of the reported issue was found to be a malfunctioning heater and ac distribution board.Replaced the malfunctioning parts and the device passed functional testing.
|
|