• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD ARCHITECT TOTAL B-HCG REAGENT KIT; SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD ARCHITECT TOTAL B-HCG REAGENT KIT; SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN Back to Search Results
Catalog Number 07K78-74
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2022
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Event Description
The customer stated that a negative architect total b-hcg result of <1.20 miu/ml was generated for a patient sample.Testing using an alternate method was positive.Retesting generated a negative architect total b-hcg result of <1.20 miu/ml and positive result using the alternate method.The architect instrument did not generate any errors.Qc was in range, and other sample results were normal.The sample did not have hemolysis, jaundice, lipemia, or fibrin.The customer was informed regarding possible causes of discrepancies between different methodologies and was instructed to have additional testing performed to confirm the patient's status.No impact to patient management was reported.
 
Manufacturer Narrative
Trending review determined no adverse trend for the issue for the product.Historical complaint review determined there is normal complaint activity for the lot number.Device history record review did not show any potential non-conformances, or deviations.Labeling was reviewed and found to adequately address the issue under review.Accuracy testing was completed using panels which mimic patient samples using an in-house retained kit stored at the recommended storage condition.All specifications were met indicating that the lot is performing acceptably.Based on the investigation, no deficiency for lot number 31141ud02 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT TOTAL B-HCG REAGENT KIT
Type of Device
SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14604929
MDR Text Key294658844
Report Number3005094123-2022-00113
Device Sequence Number1
Product Code DHA
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/18/2022
Device Catalogue Number07K78-74
Device Lot Number31141UD02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/17/2022
Initial Date FDA Received06/06/2022
Supplement Dates Manufacturer Received07/12/2022
Supplement Dates FDA Received07/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/07/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARC I2000SR INST, 03M74-02, ISR61418
Patient SexFemale
-
-