ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100R
Device Problems Loose or Intermittent Connection (1371); Insufficient Information (3190); Intermittent Communication Failure (4038)
Patient Problems Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 05/24/2022
Event Type  Injury  
Event Description
The recipient was reportedly explanted.Advanced bionics is in the process of obtaining additional information.When additional information is received, a supplemental report will be submitted.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
The recipient reportedly experienced no lock.Device testing could not be completed due to the no lock.On (b)(6) 2022, the recipient was reimplanted with another cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Additional information: section d.9.Advanced bionics received permission on behalf of the recipient to proceed with failure analysis on 11/08/2022.The explanted device was received at the company on 12/27/2022 and is currently undergoing analysis.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.This is not believed to be related to the return reason.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.This device was explanted for medical reasons.This older device configuration is not currently manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.This is not believed to be related to the return reason.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.This device was explanted for medical reasons.This older device configuration is not currently manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock could not be obtained at any spacing.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical tests performed.The residual gas analysis test was above the test limit.It is believed that the failure of this implantable cochlear stimulator (ics) device was caused by a loss of hermetic seal, which was concluded from the residual gas analysis data.This ultimately caused the device to cease functioning.This older device configuration is no longer manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
jennifer rhudy
28515 westinghouse place
valencia, CA 91355
MDR Report Key14605466
MDR Text Key293435651
Report Number3006556115-2022-00914
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/17/1999
Device Model NumberAB-5100R
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/06/2022
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Not provided
Not provided
Supplement Dates FDA Received06/23/2022
01/17/2023
04/25/2023
04/25/2023
06/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/1997
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age32 YR
Patient SexFemale
-
-