Model Number AB-5100R |
Device Problems
Loose or Intermittent Connection (1371); Insufficient Information (3190); Intermittent Communication Failure (4038)
|
Patient Problems
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 05/24/2022 |
Event Type
Injury
|
Event Description
|
The recipient was reportedly explanted.Advanced bionics is in the process of obtaining additional information.When additional information is received, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient reportedly experienced no lock.Device testing could not be completed due to the no lock.On (b)(6) 2022, the recipient was reimplanted with another cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Additional information: section d.9.Advanced bionics received permission on behalf of the recipient to proceed with failure analysis on 11/08/2022.The explanted device was received at the company on 12/27/2022 and is currently undergoing analysis.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.This is not believed to be related to the return reason.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.This device was explanted for medical reasons.This older device configuration is not currently manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.This is not believed to be related to the return reason.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.This device was explanted for medical reasons.This older device configuration is not currently manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock could not be obtained at any spacing.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical tests performed.The residual gas analysis test was above the test limit.It is believed that the failure of this implantable cochlear stimulator (ics) device was caused by a loss of hermetic seal, which was concluded from the residual gas analysis data.This ultimately caused the device to cease functioning.This older device configuration is no longer manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|