Model Number 0998-00-3023-53
Device Problem Suction Problem (2170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2022
Event Type  malfunction  
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) vacuum pump was low.No patient harm, serious injury or adverse event was reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the iabp and found that the iab fill port hose was cut with something sharp.To fix the issue, the fse removed a small piece of the hose and reattached it, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key14626519
MDR Text Key294258162
Report Number2249723-2022-01354
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/17/2022
Initial Date FDA Received06/07/2022
Supplement Dates Manufacturer Received02/28/2024
Supplement Dates FDA Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2015
Is the Device Single Use? No
Type of Device Usage Reuse