OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-390
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the fluid pump did not engage and overheated.No patient injury was reported.
 
Manufacturer Narrative
Other text: h6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device enclosure front cover/water tank cover cracked, drain/quick conn fitting corroded, line cord worn, pump faulty not engaging, came in with a 90 plate clip but the equipment is a 390.The customer stated problem was duplicated.Removed and replaced cover/water tank and reservoir gasket due to it leaking and being cracked, block assembly due to a hl 90 block being on the device when it is a hl 390, heater due to it being faulty, pump due to being faulty by not engaging, line cord due to it being worn, pole clamp due to it being worn, drain/quick connect fitting due to being corroded, front cover due to being worn, interlock block o-rings, reflux plug o-rings.The cause of the reported problem could not be determined.A dhr (device history review) was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer reported problem were found during the review of service and repair records., corrected data: e4: correction: customer reported to fda: no information.D5: correction: operator of the device: unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14631109
MDR Text Key293548650
Report Number3012307300-2022-11059
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-