|
Model Number 84592 |
Device Problems
Fracture (1260); Failure to Advance (2524)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/27/2022 |
Event Type
malfunction
|
Event Description
|
Reportable based on device analysis completed on 17may2022.It was reported that the device was fractured at the hub.The stenosed target lesion was located in the moderately tortuous and moderately calcified left prostate artery.A ngmc/fathom/021/bern/1ro/130 direxion fathom-16 system was selected for use.During the procedure, it was noted that the device was fractured at the proximal hub.The device was removed intact and the procedure was completed with another of the same device.There were no complications reported.However, device investigations revealed that the device was fractured 25.3cm from the hub.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.The device shaft was analyzed for any damage.The devices shaft showed a fracture located 25.3cm from the hub.The fracture looked to be consistent with a tensile force breaking.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.A fracture was confirmed.
|
|
Search Alerts/Recalls
|
|
|