• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAEMONETICS, INC. CELL SAVER® ELITE® PLUS AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 125ML

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAEMONETICS, INC. CELL SAVER® ELITE® PLUS AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 125ML Back to Search Results
Model Number CSE-P-125
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2022
Event Type  malfunction  
Event Description
On april 27, 2022, haemonetics was notified of a long empty error message with over estimation of rbc volume (but no waste contamination).The error message was seen on emptying of the 10th bowl and final bowl on this device utilizing the cell saver® elite processing set - 125ml in the united kingdom.There were no obvious internal cracks to the bowl reported at this time.
 
Manufacturer Narrative
On june 2, 2022, it was reported the investigation found that an inner core leak was observed on a video that was provided and cracks at the core area of the bowl were confirmed.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due to the event of the past report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER® ELITE® PLUS AUTOTRANSFUSION SYSTEM
Type of Device
CELL SAVER ELITE SET - 125ML
Manufacturer (Section D)
HAEMONETICS, INC.
125 summer street
boston MA 02110
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer Contact
brenda bruyere
125 summer street
boston, MA 02110
MDR Report Key14633348
MDR Text Key300468374
Report Number1219343-2022-00032
Device Sequence Number1
Product Code CAC
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCSE-P-125
Device Lot Number121024
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/02/2022
Initial Date FDA Received06/08/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-