On april 27, 2022, haemonetics was notified of a long empty error message with over estimation of rbc volume (but no waste contamination).The error message was seen on emptying of the 10th bowl and final bowl on this device utilizing the cell saver® elite processing set - 125ml in the united kingdom.There were no obvious internal cracks to the bowl reported at this time.
|
On june 2, 2022, it was reported the investigation found that an inner core leak was observed on a video that was provided and cracks at the core area of the bowl were confirmed.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due to the event of the past report.
|