Model Number 61000
Device Problems Mechanical Problem (1384); Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2022
Event Type  malfunction  
Manufacturer Narrative
The terumo bct technician evaluated the machine at the customer site.The iv pole was adjusted then verified to be operational.Investigation is in process.A follow up report will be provided.
 
Event Description
The customer reported that the iv pole would not stay up.There was not a patient involved at the time of the iv pole falling, therefore patient information is not reasonably known.
 
Event Description
The customer reported that the iv pole would not stay up.There was not a patient involved at the time of the iv pole falling, therefore patient information is not reasonably known.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: the terumo bct technician evaluated the machine at the customer site.The iv pole was adjusted then verified to be operational.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no other problems identified related to the reported condition.Root cause: a root cause assessment was performed for this complaint.The root cause was determined to be related to component wear and tear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA OPTIA
Type of Device
SPECTRA OPTIA APHERESIS SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key14645068
MDR Text Key301933864
Report Number1722028-2022-00189
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583610002
UDI-Public05020583610002
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number61000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/17/2022
Initial Date FDA Received06/09/2022
Supplement Dates Manufacturer Received07/19/2022
Supplement Dates FDA Received07/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/05/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-