• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VERSASCOPE INNER SHEATH; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VERSASCOPE INNER SHEATH; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Catalog Number GMS805
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Photo analysis: the device will not be returned for evaluation, but photographic evidence has been provided that appears to confirm the issue reported.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.To date it has been reported that the device will not be returned.If the device or further details are received at a later date a supplemental medwatch will be sent.(b)(4).
 
Event Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and a hysteroscope was used.The end of the device where the plastic is has been become detached.The sheath where the grasper is inserted down had a small piece missing from it.Under the scope the surgeon performed a full scope of the uterus but could not see it.As part of the procedure they flush the uterus after inspection and no foreign body (sheath broken piece) was present in the uterus.Procedure extension was no more than 15 minutes with thorough inspection and patient was fine.There were no adverse patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSASCOPE INNER SHEATH
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
CONMED
11311 concept boulevard
largo FL 33773
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key14649495
MDR Text Key301492776
Report Number2210968-2022-04446
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K972426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue NumberGMS805
Device Lot Number2020060180
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/31/2022
Initial Date FDA Received06/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/21/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-