Catalog Number 100/512/090
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that after two days of use, the white cuff broke.Additional information received indicated that medical intervention was required.The tracheostomy tube was changed at the patient's home by the doctor.No patient injury was reported.
 
Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Three (3) product sample photos were received for evaluation.Visual inspection was conducted and could be observed of a top view of the packaging from each product reported for customer; however, none of them show the condition of the cuff and it is not possible confirm the failure mode reported by customer.No functional testing could be performed as no samples were received.No root cause could be determined since the complaint was not confirmed due to photos provided does not show the failure mode reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUE LINE TRACHEOSTOMY TUBES
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14655742
MDR Text Key294121257
Report Number3012307300-2022-11555
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeGR
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/21/2023
Device Catalogue Number100/512/090
Device Lot Number3731681
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/12/2022
Initial Date FDA Received06/10/2022
Supplement Dates Manufacturer Received08/05/2022
Supplement Dates FDA Received09/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-