BOSTON SCIENTIFIC NEUROMODULATION SUPERION INDIRECT DECOMPRESSION SYSTEM; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE Back to Search Results
Model Number 101-9812
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Pain (1994)
Event Date 05/20/2022
Event Type  Injury  
Manufacturer Narrative
Event date: approximated based on the date the manufacturer became aware of the event.
 
Event Description
It was reported the patient experienced a return of pre-implant pain approximately 4 weeks after being implanted with the superion indirect decompression spacer.Imaging taken in the field confirmed the spacer had migrated anteroposterior from between the l4/l5 lumbar vertebrae.The physician assessed the cause of the spacer having become dislodged was due to the patient being active too soon after implantation, not allowing enough time to heal.The patient underwent a procedure where the spacer was removed and did well post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERION INDIRECT DECOMPRESSION SYSTEM
Type of Device
PROSTHESIS, SPINOUS PROCESS SPACER/PLATE
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
VERTIFLEX INC.
2714 loker ave. west
carlsbad CA 92010
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key14660115
MDR Text Key293844094
Report Number3006630150-2022-02817
Device Sequence Number1
Product Code NQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number101-9812
Device Catalogue Number101-9812
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-