Catalog Number 74120152
Device Problem Biocompatibility (2886)
Patient Problems Pain (1994); Metal Related Pathology (4530)
Event Date 05/17/2022
Event Type  Injury  
Event Description
It was reported that, after a bhr system was implanted on (b)(6) 2005 due to arthrosis, the patient experienced metallosis and pain.This adverse event was treated with a revision surgery on (b)(6) 2022.Patient's current health status is unknown.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
H3, h6: it was reported that, after a bhr system was implanted, the patient experienced metallosis and pain.This adverse event was treated with a revision surgery.Patient's current health status is unknown.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.Without definitive batch numbers, a complete review of the historical complaints data cannot be performed for the devices concerned.A review of historical complaints data was performed using the part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints were identified for the femoral head and the acetabular cup.However, as the devices are no longer sold, no action is to be taken.Without definitive lot numbers a complete complaint history review cannot be performed for the devices involved.Should further details become available at a later date then the task will be reopened and completed.As no device batch numbers were provided for investigation, manufacturing record review, device labelling / ifu review could not be performed.If more information is received, this investigation will be reopened.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.The available medical documents were reviewed.The provided ¿surgical report¿ was reviewed; however, it not contain details of the surgical procedure.Without supporting medical documentation, a thorough medical assessment cannot be performed.In the event additional medical/clinical records are received, the clinical task may be re-opened, and a thorough assessment will be rendered at that time.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 52MM W/ IMPTR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14670221
MDR Text Key294858195
Report Number3005975929-2022-00397
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010502575
UDI-Public3596010502575
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number74120152
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/30/2022
Initial Date FDA Received06/11/2022
Supplement Dates Manufacturer Received09/14/2022
Supplement Dates FDA Received09/24/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2746-2015
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age51 YR
-
-