AVANOS MEDICAL INC. CORTRAK 2 ENTERAL ACCESS SYSTEM; DH CORTRAK (EAS) Back to Search Results
Model Number 20-0950
Device Problem Malposition of Device (2616)
Patient Problem Pneumothorax (2012)
Event Date 05/11/2022
Event Type  Injury  
Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.Root cause could not be determined.All information reasonably known as of 10 jun 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.Device not returned.
 
Event Description
Avanos medical inc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the first of two reports.Refer to 3011270181-2022-00043 for the second report.A pneumothorax was reported.No additional information was provided concerning the patient¿s current condition.
 
Event Description
Additional information received on 16jun2022 reported that, to the best of the reporter¿s knowledge that the patient was injured but did recover.Additionally, the policy was to order an x-ray after every placement.
 
Manufacturer Narrative
All information reasonably known as of 30 jun 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Manufacturer Narrative
H6: investigation findings (appropriate term/code not available): user, training issue.The actual sample from the reported event was not returned for evaluation; however, a potential misplacement was visible on the mu screen (via the provided tracing(s)), along with confirmation from the reporter that the user was not trained at bedside and had not completed the minimum number of supervised placements; the root cause was determined to be "user: incorrect use." the device history record for lot 1204001 was reviewed and the product was produced according to product specifications.All information reasonably known as of 29 jul 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Manufacturer Narrative
All information reasonably known as of 12 apr 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORTRAK 2 ENTERAL ACCESS SYSTEM
Type of Device
DH CORTRAK (EAS)
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
MINNETRONIX, INC.
1635 energy park dr.
saint paul MN 55108
Manufacturer Contact
ujjal chakravartty
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key14672579
MDR Text Key294876117
Report Number3011270181-2022-00042
Device Sequence Number1
Product Code KNT
UDI-Device Identifier00350770472010
UDI-Public00350770472010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K220588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup,Followup,Followup
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Model Number20-0950
Device Catalogue NumberN/A
Device Lot Number1204001
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
RU RECEIVER SN (B)(6)
Patient Outcome(s) Other; Life Threatening;
Patient SexFemale
-
-