Model Number WB91051W
Device Problems Sparking (2595); Output Problem (3005); Noise, Audible (3273); Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/16/2022
Event Type  malfunction  
Event Description
Olympus was informed that during preparartion for use for an unspecified procedure the hf generator unit made a loud bang and visible sparks could be seen from the side.The intended procedure was completed using another similar device and there was no report about an adverse event or patient injury.When the hf generator was examined at the olympus regional repair center it was found that it issued error message e433.
 
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation but to an olympus regional repair center (rrc) in the united kingdom (returned to rrc on june 6, 2022).The evaluation at the rrc confirmed the occurrence of error message e433 which is triggered by the generator¿s safety system.In case of critical errors, the safety system will not permit any further use of the generator until the error is rectified.In the case at hand, this was caused by a defective high voltage power supply (hvps) board.Thus, this event/incident was attributed to component failure.The bang and sparks reported by the customer were most likely a side effect of this failure.A manufacturing and quality control review was performed for the affected serial number of the hf generator without showing any abnormalities.The case will be closed on olympus side with no further actions, but the reported event/incident will be recorded for trending and surveillance purposes.Furthermore, the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF UNIT "ESG-400"
Type of Device
HF-GENERATORS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
rheinstrasse 8
teltow
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key14673106
MDR Text Key302182858
Report Number9610773-2022-00213
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04042761076838
UDI-Public04042761076838
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K203682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWB91051W
Device Catalogue NumberWB91051W
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/04/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-