Model Number 4542 |
Device Problems
Failure to Capture (1081); Fracture (1260); Impedance Problem (2950); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/12/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that this left ventricular (lv) lead was suspected to have fractured.The patient arrived at the hospital for a routine follow-up without capture on their right ventricular (rv) lead.The device was programmed with the configuration lv tip to lv ring, but this resulted in loss of capture and out-of-range impedance measurements.The configuration was changed to lv ring to can with normal pacing threshold and good and stable impedance.The plan for this patient is continued routine follow-ups.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
This device remains implanted; therefore, technical analysis cannot be conducted.Without a returned device it is not possible to definitively confirm how the device may have contributed to the complaint incident.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this left ventricular (lv) lead was suspected to have fractured at the tip, resulting in out-of-range impedances and loss of capture.The issue was resolved through electronic repositioning of the lead (programming the lead tip out of the sensing/pacing configuration).The plan for this patient is continued routine follow-ups.No adverse patient effects were reported.
|
|
Search Alerts/Recalls
|