• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE; DERMATOME, PNEUMATICALLY-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE; DERMATOME, PNEUMATICALLY-POWERED Back to Search Results
Model Number N/A
Device Problems Failure to Cut (2587); Physical Resistance/Sticking (4012)
Patient Problem Abrasion (1689)
Event Date 05/18/2022
Event Type  Injury  
Event Description
It was reported that the device gets stuck (chops) and goes unevenly.The event timing was during surgery.There was a 30 minute delay and an additional graft was needed.
 
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.
 
Event Description
No additional event information available.
 
Manufacturer Narrative
Review of the most recent repair record determined the repair site was unable to duplicate the reported issue.The calibration was out at the zero reading, the machined head and screw were damaged, and the reciprocating arm were worn.The machined head, screws, reciprocating arm, and bearings were replaced and resolved the reported issue.It was noted that this unit was manufactured in november of 2017 and has not been returned for preventative maintenance.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMATOME AN HANDPIECE
Type of Device
DERMATOME, PNEUMATICALLY-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key14675066
MDR Text Key293843129
Report Number0001526350-2022-00560
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024464834
UDI-Public(01)00889024464834(11)171103(10)63817639
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number88710100
Device Lot Number63817639
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/06/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-