Model Number N/A |
Device Problems
Failure to Cut (2587); Physical Resistance/Sticking (4012)
|
Patient Problem
Abrasion (1689)
|
Event Date 05/18/2022 |
Event Type
Injury
|
Event Description
|
It was reported that the device gets stuck (chops) and goes unevenly.The event timing was during surgery.There was a 30 minute delay and an additional graft was needed.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.
|
|
Event Description
|
No additional event information available.
|
|
Manufacturer Narrative
|
Review of the most recent repair record determined the repair site was unable to duplicate the reported issue.The calibration was out at the zero reading, the machined head and screw were damaged, and the reciprocating arm were worn.The machined head, screws, reciprocating arm, and bearings were replaced and resolved the reported issue.It was noted that this unit was manufactured in november of 2017 and has not been returned for preventative maintenance.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional event information available.
|
|