• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD ENDO CLIP III; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD ENDO CLIP III; CLIP, IMPLANTABLE Back to Search Results
Model Number 176630
Device Problems Difficult to Remove (1528); Difficult to Open or Close (2921); Fail-Safe Problem (2936)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/20/2022
Event Type  malfunction  
Event Description
According to the reporter, during the laparoscopic cholecystectomy, while clipping the blood vessel and after loading, the handle became unable to be squeezed.The jaws could not be closed. it was also stated that it could not be removed from the 5 millimeter port, so it was removed together with the port.Another product was used to resolve the issue.There was no patient injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted a jaw leg was out of position, the cam on the jaw leg was not aligned with the driver.It was reported that the jaws of the reload did not close at all, there was difficulty removing the device from inside the port and the safety interlock deployed prior to all staples or clips firing.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The issue can occur if one of the jaws legs gets forcefully pulled outward away from the center line of the shaft.The noted jaw cam disengagement impedes functionality of the jaws, possibly resulting in clip malformation or difficulty removing the instrument from a trocar.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: avoid excessive twisting of the jaws or tissue manipulation when firing the instrument.Deflecting the jaws and or the shaft during firing may result in an improperly formed clip and possible bleeding and or leakage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP III
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
5920 longbow drive
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key14676815
MDR Text Key293865765
Report Number9612501-2022-00959
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521057852
UDI-Public10884521057852
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K100242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number176630
Device Catalogue Number176630
Device Lot NumberJ1M1003Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/27/2022
Initial Date FDA Received06/13/2022
Supplement Dates Manufacturer Received08/19/2022
Supplement Dates FDA Received08/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-