Model Number 71992-01 |
Device Problems
Device Displays Incorrect Message (2591); Insufficient Information (3190)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 06/02/2022 |
Event Type
Injury
|
Event Description
|
A customer reported an error message of ¿replace sensor¿ was received on the same day of applying the adc device.As a result, the customer reported receiving third-party medical treatment however, no further details were reported.Attempts to gather additional information have been unsuccessful thus far as the customer refused to answer further questions.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The date the incident occurred is unknown.The date entered is the date abbott diabetes care became aware of the event.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer reported an error message of ¿replace sensor¿ was received on the same day of applying the adc device.As a result, the customer reported receiving third-party medical treatment however, no further details were reported.Attempts to gather additional information have been unsuccessful thus far as the customer refused to answer further questions.There was no report of death or permanent injury associated with this event.
|
|
Search Alerts/Recalls
|